What are the Limitations of DLS?

Published in Particle Size Analysis 3 mins read

Dynamic Light Scattering (DLS) is a powerful technique used to determine the size and size distribution of particles in a solution. While DLS offers numerous advantages, it also has some limitations:

1. Sensitivity to Concentration

  • Issue: DLS is highly sensitive to particle concentration. Too high a concentration can lead to multiple scattering events, making it difficult to accurately interpret the data.
  • Solution: Dilute the sample to ensure single scattering events dominate.
  • Example: If you're analyzing a concentrated protein solution, diluting it will provide more accurate size measurements.

2. Polydispersity

  • Issue: DLS struggles to accurately analyze samples with a wide range of particle sizes (polydisperse samples). This is because the signal from larger particles can overwhelm the signal from smaller particles.
  • Solution: Use techniques like field-flow fractionation (FFF) or size exclusion chromatography (SEC) to separate the particles by size before DLS analysis.
  • Example: A mixture of nanoparticles with sizes ranging from 10nm to 100nm might be difficult to analyze accurately with DLS alone.

3. Shape Effects

  • Issue: DLS primarily measures the hydrodynamic radius, which is the radius of a sphere that diffuses at the same rate as the particle. This means the shape of the particle can affect the DLS measurement.
  • Solution: Use other techniques, such as electron microscopy, to determine the actual shape of the particle.
  • Example: A long, rod-shaped particle will have a different hydrodynamic radius than a spherical particle of the same volume.

4. Limitations in Analysis

  • Issue: DLS data analysis can be complex, requiring specialized software and knowledge to interpret the results.
  • Solution: Invest in training or consult with experts to ensure accurate data analysis.
  • Example: Determining the true size distribution of a sample requires understanding the different parameters used in the DLS analysis software.

5. Presence of Multiple Species

  • Issue: DLS is less effective at analyzing samples containing multiple different particle types. The signal from the different species can overlap, making it challenging to distinguish between them.
  • Solution: Consider using other separation techniques, such as chromatography, before DLS analysis.
  • Example: A sample containing both proteins and liposomes might be difficult to analyze accurately with DLS alone.

6. Sample Requirements

  • Issue: DLS requires a sample that is stable and homogenous. Samples that are too viscous or contain large aggregates can interfere with the measurement.
  • Solution: Ensure the sample is properly prepared and filtered before analysis.
  • Example: A sample with large, insoluble particles will not be suitable for DLS analysis.

Conclusion

DLS is a valuable tool for particle sizing, but it is important to be aware of its limitations. By understanding these limitations and taking appropriate steps to mitigate them, you can obtain accurate and reliable results.